Instructions for use Zeus Infinity Empowered SW 2.n 259
Fault – Cause – Remedy
Warning 180 Apnea pressure No automatic ventilation: No
press. breath detected for
30 sec. No spont/ass. venti-
lation: No breath detected by
press. for 60 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Caution 250 Apnea pressure No automatic ventilation: No
press. breath detected for
15 sec. No spont/ass. venti-
lation: No breath detected by
press. for 30 sec.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Caution 025 Apnea ventilation No Pressure Support cycle
triggered within time limit.
Check patient. Use manda-
tory ventilation. Use lower
flow trigger level.
Note 050 ART check zero Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Zero the transducer.
Caution 090 ART D high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 090 ART D low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Note 025 ART hardware fail Invasive blood pressure
hardware failure.
Check hardware and re-
place. Call DrägerService.
Caution 090 ART M high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy