Startup
96
Instructions for use Zeus Infinity Empowered SW 2.n
Self test
Preparing the self test
1 Place the Y-piece (A) with the filter firmly onto
the cone (self-test adapter).
2 Connect sample line to the filter.
3 Set APL valve to 20 mbar.
4 Switch off manual O
2 supply.
Starting the self test manually
The self test can only be started from the system
standby screen.
1 Touch the Device test button (A).
The Test details dialog window will be opened.
2 Touch the Self test button (B).
3 Touch the Name of tester button (C).
4 Enter a new name or select a name from an ex-
isting list of the last six testers. Names can also
be deleted from the list.
5 The Start (D) button automatically turns yellow.
Confirm with rotary knob.
WARNING
Risk of patient injury
The self test must never be performed when a
patient is connected!
CAUTION
Increased occupational exposure
When filling during the warm-up phase of the
DIVA metering module, anesthetic agent may es-
cape.
Only fill the DIVA metering module after the warm-
up phase.