Assembly and preparation
82
Instructions for use Zeus Infinity Empowered SW 2.n
Finishing the filling process
1 Swivel the anesthetic agent bottle (A) down-
wards.
2 Check the filling level with the sight glass (B).
The DIVA metering module must be in an up-
right position.
3 Swivel the lever (C).
4 Pull out the filling adapter (D).
5 Insert the locking bolt (E) and press in com-
pletely.
6 Close the lever (C).
Otherwise, fresh gas and anesthetic agent va-
por could escape after switching on and trans-
porting the DIVA metering module.
If the lever cannot be closed completely, disengage
the lever and depress the locking bolt completely,
otherwise the locking bolt will not seal and the gas-
ket seal may be damaged.
7 Unscrew the filling adapter and close the anes-
thetic agent bottle with the cap.
To use anesthetic agents in a safe and environmen-
tally-friendly manner, please follow the following in-
structions:
058059
A
4
B
C
D
E
C
NOTE
Do not store any anesthetic agent bottles with the
filling adapter unscrewed! Anesthetic agent will be
released!
NOTE
Close the anesthetic agent bottle, even if it has
been emptied. Allow remaining anesthetic agent
in the filling adapter and anesthetic agent bottle to
evaporate under an extractor hood, otherwise an-
esthetic agent vapor will be released.