Technical data
420
Instructions for use Zeus Infinity Empowered SW 2.n
Classification Caution
Alarm On/Off Off: No alarms generated
Automatic alarm limits Current measured value ±2 mm
Response time to changes in heart rate
Change in heart rate from 90 to 120 bpm Range: 3.4 to 7.1 s
Average: 5.3 s
Change in heart rate from 90 to 40 bpm Range: 6.3 to 8.6 s
Average: 7.4 s
ECG (Cont.)
Non-invasive blood pressure measurement (NIBP)
Parameter display Systolic, diastolic, mean
Measuring principle Oscillometry
Modes of operation Manual (single measurement), continuous (5 min-
utes), interval or stasis
Interval times 1, 2, 3, 5, 10, 15, 30 min
Measuring ranges (adults), 270 mmHg Heart rate: 40 to 240 bpm
NIBP (systolic): 30 to 250 mmHg
NIBP (mean): 20 to 230 mmHg
NIBP (diastolic): 10 to 210 mmHg
Measuring ranges (children), 180 mmHg Heart rate: 40 to 240 bpm
NIBP (systolic): 30 to 170 mmHg
NIBP (mean): 20 to 150 mmHg
NIBP (diastolic): 10 to 130 mmHg
Measuring ranges (neonates), 140 mmHg Heart rate: 40 to 240 bpm
NIBP (systolic): 30 to 130 mmHg
NIBP (mean): 20 to 110 mmHg
NIBP (diastolic): 10 to 100 mmHg
Connections Quick-release hose connector with single airway