Instructions for use Zeus Infinity Empowered SW 2.n 421
Technical data
Default inflation pressure Adults: 160 mmHg ±10 mmHg
Children: 120 mmHg ±10 mmHg
Neonates: 110 mmHg ±10 mmHg
Inflation pressure after a valid measurement
(±10 mmHg)
Adult: Previous systolic pressure +25 mmHg
Children: Previous systolic pressure +25 mmHg
Neonatal: Previous systolic pressure +30 mmHg
Maximum cuff pressure Adults: 265 mmHg ±5 mmHg
Children: 180 mmHg ±10 mmHg
Neonates: 142 mmHg ±10 mmHg
Minimum cuff pressure necessary for measure-
ment
Adults: 110 mmHg ±10 mmHg
Children: 90 mmHg ±10 mmHg
Neonates: 70 mmHg ±10 mmHg
Maximum inflation time Adults: 2 min ±1 s
Children: 2 min ±1 s
Neonates: 90 s ±1 s (60 s according to French law)
Software safety cut-off Adults: 273 mmHg ±3 mmHg
Children: 215 mmHg ±3 mmHg
Neonates: 153 mmHg ±3 mmHg
Hardware safety cut-off Adults and children: 300 mmHg ±30 mmHg
Neonates: 170 mmHg ±20 mmHg
Static cuff pressure accuracy ±3 mmHg
Calibration range Adults/children: 10 to 260 mmHg ±3 mmHg
Neonates: 10 to 150 mmHg ±3 mmHg
Electrical specifications Type CF as per IEC 60601-1
Defibrillation protection according to IEC 60601-2-30
Non-invasive blood pressure measurement (NIBP) (Cont.)