Instructions for use Zeus Infinity Empowered SW 2.n 167
Alarms
Alarm recorder (optional with hemodynamic and Infinity network)
Select alarms for which the alarm recorder auto-
matically starts when they occur.
z Touch the Alarm recorder tab (A).
B Recorder triggers
Selection of alarms that start the alarm recorder
C Recorder config.
Specification of the alarm recorder parameters
Calling up Recorder triggers:
1 Touch the Recorder triggers tab (B).
2 The arrow keys ( ) are used to add or re-
move available Trigger alarms (D) for the
alarms Active triggers (E).
3 Use the Apply changes button (F) to confirm
the changes.
Calling up Recorder config.:
1 Touch the Recorder config. tab (C).
The following settings can be configured:
G Record alarms: OFF/ON
H Recorder speed
I Recorder delay
J Duration
K Waveform 1
L Waveform 2
M Alarm waveform: OFF/ON
2 Adjust settings.
3 Use the rotary knob to set and confirm the
value.
4 Use the Apply changes button (N) to confirm
the changes.
WARNING
National and international standards specify
minimum monitoring with alarm functions.
These regulations may not be complied with if
the alarm function of the monitoring parame-
ter etCO
2 is disabled!