Instructions for use Zeus Infinity Empowered SW 2.n 327
Configuration
Profiles Default
The Default property defines the default setting
which is loaded when the device is switched on (not
with Views).
Only profiles that are assigned to a profile key can
be defined as the default.
The Manage setups > Profiles page opens.
If a specific start setting is required:
1 Use the rotary knob to select the required profile
from the Profile name list.
2 Touch the Default button (A).
3 Enter access code and confirm with rotary
knob.
The Default button is now dark green. This profile
is preset after switching on the device and when
calling up new patient data (default setting).
Protecting profiles/views
With the Protect property, the profile or view setting
cannot be renamed, deleted or saved.
The Manage setups > Profiles > or Views page
opens.
1 Touch the Profiles or Views tab.
2 Use the rotary knob to select the profile or the
view to be protected from the Profile name list
or the View name list.
3 Touch the Protect button (A).
The button turns dark green and the dialog window
for the access code opens.
4 Enter the access code and confirm with rotary
knob.
Release profiles/views
Unlocking protected profiles or views:
1 Select the profile or view.
2 Touch the Protect button.
3 Enter the access code and confirm with rotary
knob.
The required setting is then unlocked.