Technical data
426
Instructions for use Zeus Infinity Empowered SW 2.n
Power of radiation sources Nellcor: Red: 3 mW (max.), Infrared: 4 mW (max.)
Masimo: Red: 0.9 mW (max.), Infrared: 0.9 mW
(max.)
Note: LED is voltage-limited by a hardware mecha-
nism.
Electrical specifications Type CF as per IEC 60601-1
Defibrillation protection according to IEC 60601-1A2
Temperature
Parameter display Absolute temperature
Measuring range –5 °C to 50 °C (absolute)
Resolution ±0.1 °C
Accuracy ±0.1 °C (absolute)
Average response time <2.5 seconds
Sensor accuracy ±0.1 °C; 0 °C to 50 °C
Electrical specifications Type CF as per IEC 60601-1
Defibrillation protection according to IEC 60601-1A2
Neuromuscular Transmission Monitoring NMT
Stimulation modes Single Twitch (Single), Train of Four (TOF), Post
Tetanic Count (PTC)
Range for stimulation current Auto: Display only; 60 mA max. up to 6.6 kΩ; based
on the reference pulse at start
Man: 5 to 60 mA in 5 mA increments
Pulse width 100, 200 or 300 µs
Measuring intervals None, 1 s, 10 s, 20 s, 1 min, 5 min, 15 min, 30 min
Pulse oximetry (SpO2) (Cont.)