Instructions for use Zeus Infinity Empowered SW 2.n 85
Assembly and preparation
3 Screw the locking cap (C) off the filling device
slowly so that any pressure which has built up in
the DIVA metering module can escape slowly.
4 Insert the anesthetic agent bottle with the bars
into the respective grooves (D) of the filling
opening. Only use anesthetic agent bottles with
suitable bars!
z Press the anesthetic agent bottle up to the stop
into the filling opening and keep it pressed.
z Avoid excessive force and tilting of the anes-
thetic agent bottle!
z Watch the filling level in the sight glass.
The supply will be interrupted automatically
when the maximum mark is reached.
Finishing the filling process
1 Reduce the pressure in the anesthetic agent
bottle (A) and slowly pull out the anesthetic
agent bottle.
2 Check the filling level in the sight glass (B) –
DIVA metering module must stand vertically
during checking.
z Screw the cap onto the bottle adapter.
z Always keep the anesthetic agent bottle closed!
061
C
D
062
CAUTION
Increased occupational exposure
Firmly screw on the locking cap. Otherwise, anes-
thetic agent can escape.
B
A