Operation
146
Instructions for use Zeus Infinity Empowered SW 2.n
Extreme settings on the syringe pump, which lead
to a delivery rate of >1200 mL/h, may under certain
circumstances be a departure from the set range of
the infusion therapy control or bolus therapy
control. This is indicated on the therapy control by
> max.
z Check concentration and weight setting, and
adjust if necessary. Select suitable metering.
Syringe change
If the content of a syringe is used up completely
during the course of an infusion, it can be replaced
with another syringe filled with the same drug when
the syringe pump is on. The simultaneous replace-
ment of syringe and drug is described on page 148.
Stopping the infusion
1 Touch the Start infusion button (A).
The therapy control (B) for the infusion rate is dis-
played with the preset value.
2 Set 0 with the rotary knob and confirm.
Infusion is stopped.
Stopping the infusion in the Infusion settings
dialog window
z Touch the Off button (C) and confirm with rotary
knob.
If an infusion is restarted, the infusion rate value
last set is displayed on the therapy control.
Administering a bolus
A bolus can be started after switching on a Module
DPS syringe pump or during an infusion. The initial
bolus is displayed in red.
After the second bolus and after transporting the
Module DPS to Zeus IE, the maintenance bolus is
loaded. If the bolus is changed for the maintenance
bolus, that dose is loaded and administered for the
following maintenance boluses.
Checking patient data/weight
Before starting the initial bolus, the correct patient
weight must be entered as when starting a first in-
fusion, see page 144.
Starting a bolus
z Touch the Start bolus button (A).
Or
z In the Infusion settings (see page 142) dialog
window, touch the On button for the Bolus.