Instructions for use Zeus Infinity Empowered SW 2.n 75
Assembly and preparation
Sight glasses
The DIVA metering modules have a sight glass with
marks for minimum and maximum filling levels. A
third mark in the lower section indicates that a en-
tire anesthetic agent bottle (250 mL) can be refilled.
Filling devices
The following filling devices are possible for the dif-
ferent anesthetic agents:
Fill the DIVA metering modules
If the brand name is different, ensure correct as-
signment, e.g., by the color coding of the DIVA me-
tering module and anesthetic agent bottle.
Standardized color coding in accordance with
ISO 5359 / ISO 32 / ISO 5360 is used to identify an-
esthetic agents and medical gases.
Anesthetic agents from different manufacturers
with different brand names can be metered in the
DIVA metering module from a technical viewpoint
individually or mixed if they are equal in composi-
tion as well as their physical and chemical features
and if they are approved for pharmaceutical use.
Close the drain plug before filling!
CAUTION
Risk of injury
Place DIVA metering modules only on firm, hori-
zontal surfaces. The DIVA metering module can
be damaged when dropped.
004
Isoflurane Safety filling system/Dräger Fill
Sevoflurane Quik Fil
Safety filling system/Dräger Fill
Desflurane Desflurane filling system
250 mL
CAUTION
Hazardous to health
Only fill DIVA metering modules in areas with ad-
equate ventilation.
Do not spill any anesthetic agents and do not in-
hale anesthetic vapor.
NOTE
Observe the expiration date and storage temper-
ature of the anesthetic agent.
WARNING
Risk of patient injury
Make sure the name of the anesthetic agent
and the color coding on the DIVA metering
module and anesthetic agent bottle agree.
Otherwise, hazardous anesthetic agent mix-
tures may develop or mistakes can be made.
WARNING
Risk of patient injury
Do not allow any detergents or substances not
intended for use to enter the filling system.
The anesthetic agent may be contaminated.