Alarms
180
Instructions for use Zeus Infinity Empowered SW 2.n
Patient-related
Alarm Setting range Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1
HR (1/min) 25 to 300 120 135 150 180
HR (1/min) 20 to 295 45 50 60 80
Brady Limit 20 to 100; Off – – 50 50 60
PVC (1/min) 1 to 50 50 50 50 50 –
ST (mm) –9.9 to 10 2
ST (mm) –10 to 9.9 –2
ART S (mmHg) 0 to 300 160 120 80
ART S (mmHg) –5 to 295 90 50
ART D (mmHg) 0 to 300 80 80 60
ART D (mmHg) –5 to 295 50 30 25
ART M (mmHg) 0 to 300 125 90 70
ART M (mmHg) –5 to 295 60 50 40
LV S (mmHg) 0 to 300 160
LV S (mmHg) –5 to 295 75
LV D (mmHg) 0 to 300 25
LV D (mmHg) –5 to 295 5
LV M (mmHg) 0 to 300 80
LV M (mmHg) –5 to 295 40
GP1 S, GP2 S (mmHg) 0 to 300 160 120
GP1 S, GP2 S (mmHg) –5 to 295 90 50
GP1 D, GP2 D (mmHg) 0 to 300 110 80
GP1 D, GP2 D (mmHg) –5 to 295 50 35
GP1 M, GP2 M (mmHg) 0 to 300 125 85
GP1 M, GP2 M (mmHg) –5 to 295 60 40
PA S (mmHg) 0 to 120 35
PA S (mmHg) –5 to 115 10
PA D (mmHg) 0 to 120 15
PA D (mmHg) –5 to 115 5
PA M (mmHg) 0 to 120 20
PA M (mmHg) –5 to 115 10
RV S (mmHg) 0 to 120 35