Instructions for use Zeus Infinity Empowered SW 2.n 181
Alarms
RV S (mmHg) –5 to 115 10
RV D (mmHg) 0 to 120 15
RV D (mmHg) –5 to 115 5
RV M (mmHg) 0 to 120 20
RV M (mmHg) –5 to 115 10
CVP M (mmHg) 0 to 120 20
CVP M (mmHg) –5 to 115 0
LA M (mmHg) 0 to 120 20
LA M (mmHg) –5 to 115 0
RA M (mmHg) 0 to 120 15
RA M (mmHg) –5 to 115 5
ICP M (mmHg) –20 to 120 20
ICP M (mmHg) –25 to 115 5
CPP M (mmHg) –20 to 300 100
CPP M (mmHg) –25 to 295 70
SpO
2 71 to 100 100 95
SpO
2 70 to 99 90 85
NIBP S (mmHg) Neo: 15 to 130
Ped: 15 to 170
Adult: 15 to 250
160 120 80
NIBP S (mmHg) Neo: 10 to 125
Ped: 10 to 165
Adult: 10 to 245
90 50 50
NIBP D (mmHg) Neo: 15 to 130
Ped: 15 to 170
Adult: 15 to 250
80 80 60
NIBP D (mmHg) Neo: 10 to 125
Ped: 10 to 165
Adult: 10 to 245
50 30 25
NIBP M (mmHg) Neo: 15 to 130
Ped: 15 to 170
Adult: 15 to 250
125 90 70
NIBP M (mmHg) Neo: 10 to 125
Ped: 10 to 165
Adult: 10 to 245
60 50 40
T1/T2 (°C) –4.9 to 50 39
Alarm Setting range Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1