Instructions for use Zeus Infinity Empowered SW 2.n 177
Alarms
Suppressing the alarm tone
The alarm tone can be suppressed for maximum
2 minutes.
z Press the Audio paused (A) button.
The audible alarm of all active alarms is sup-
pressed for 2 minutes.
In the header bar, the symbol (B) with the re-
maining silence time (minutes:seconds) appears.
If the Audio paused key is pressed a second time
during this time, the All symbol is symbol ap-
pears. All audible alarms which appear in the next
2 minutes are suppressed. The visual signal on the
alarm bar disappears.
If there are no active alarms before pressing the
Audio paused key, then all alarms which appear in
the next 2 minutes are suppressed (preventive si-
lence function).
Switching the audible alarm on again
z Press the Audio paused key (A) again.
The information field in the header bar is empty.
Exceptions
The following alarms immediately end the silence
time:
050469
O2
Air N2OO2 Air N2O
A
1:35
1:35
B
Alarm message
Air supply fail
Air+N
2O delivery cutoff
Anesth. Ventilator Off
Apnea
AW press. protect. fail
AW pressure high
AW pressure negative
Battery low
Battery very low
Cont. AW pressure
DIVA module 1 failure
DIVA module 2 failure
DIVA protect fail
DIVA+mixer failure
Fresh gas intern/extern?
Infusion line 1 blocked
Infusion line 2 blocked
Infusion line 3 blocked
Infusion line 4 blocked
Insp. desflurane high
Insp. enflurane high
Insp. halothane high
Insp. isoflurane high
Insp. oxygen low
Insp. sevoflurane high