Instructions for use Zeus Infinity Empowered SW 2.n 31
Overview
Connections, power supply and addi-
tional sockets
View with doors open, from above:
A Mains power supply
Secure the plug with screws!
B Automatic fuse
Can be reset after tripping.
C Equipotential connection
D Auxiliary sockets for auxiliary devices
Electronically fused, can be reset after the fuse
has tripped.
170
NOTE
Auxiliary sockets are not supplied by the uninter-
ruptible power supply UPS and are not supplied
with power in the event of a mains power failure!
WARNING
Do not connect HF surgery devices to the aux-
iliary sockets because interference can impair
Zeus IE functionality!
A