Instructions for use Zeus Infinity Empowered SW 2.n 179
Alarms
Set ranges of the patient monitoring alarm limits
Device-related
Alarm Setting range Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1
etCO
2 (Vol.%) 0.1 to 9.9; Off 7
etCO
2 (Vol.%) Off; 0 to 9.8 3
inCO
2 (Vol.%) 0 to 1.4; Off 1
FiO
2 (Vol.%) 19 to 100; Off Off 90
FiO
2 (Vol.%) 18 to 99 20
inHal (Vol.%) 0.1 to 7.0 2.1
inHal (Vol.%) Off; 0 to 6.9 Off
inIso (Vol.%) 0.1 to 7.0 3.1
inIso (Vol.%) Off; 0 to 6.9 Off
inEnf (Vol.%) 0.1 to 9.9 4.3
inEnf (Vol.%) Off; 0 to 9.8 Off
inDes (Vol.%) 0.1 to 21.9 17.5
inDes (Vol.%) Off; 0 to 21.8 Off
inSev (Vol.%) 0.1 to 9.9 6 7 8
inSev (Vol.%) Off; 0 to 9.8 Off
N
2O (Vol.%) 1 to 82 82
N
2O (Vol.%) 0 to 81; Off Off
MVe (L/min) 0.1 to 39.0; Off Off
MVe (L/min) 0 to 38.9 4 0.8 0.2
Paw (mbar) 5 to 100 40 25 20
V
Ti (L/0.01) 0.02 to 2.0; Off 1.2 0.5 0.1
Brody Factor 1.5 to 5.0, Off 2.5