Instructions for use Zeus Infinity Empowered SW 2.n 359
Cleaning, disinfection, and sterilization
Note on removable accessories, con-
sumable parts, and attached parts
Consult the appropriate instructions for use:
Accessories and consumable parts
– Flow sensors
– CLIC adapter
– CLIC absorber
– Breathing hoses
– Filter
– Breathing bag
–Masks
– Water trap
– DIVA modules
Attached parts
– Endotracheal suction
– Hinged arms
–Monitors
– Sensors and cables
– IT systems
– Workstation light
Disposable articles without instructions for use
– Soda lime dust filter
– Sample line
CAUTION
Risk of injury to patients due to failure of accesso-
ries
Disposable items have been designed, tested,
and manufactured exclusively for single use. Re-
use, reprocessing or sterilization can result in fail-
ure of the accessory and injury to the patient.
Do not reuse, reprocess or sterilize disposable
items.
WARNING
Risk of infection
Used sample lines may be infectious due to
the breathing gases that passed through
them.
Exchange the sample line regularly, see table
"Semicritical medical devices".
WARNING
Material damage due to disinfectants
When the sample line is disinfected and resi-
dues of the agent remain in the sample line,
these residues can get into the water trap and
the gas measurement module later. This may
result in faulty measurements.
Sample lines are disposable items and may
not be disinfected.