Instructions for use Zeus Infinity Empowered SW 2.n 429
Technical data
Essential performance
The essential performance features comprise:
General
– Supply of the anesthesia workstation with O
2
If the O2 supply (central gas supply or gas cyl-
inder) fails, an alarm is issued.
– Supply of the patient with adequately oxygen-
ated breathing gas
If the breathing gas contains insufficient levels
of O
2, an alarm is issued.
– Patients are not supplied with excessively high
anesthetic gas concentrations
If excessively high anesthetic gas concentra-
tions are delivered, an alarm is issued.
– Monitoring of the airway pressure and the expir-
atory minute volume
Alarms are issued depending on the set alarm
limits.
Gas measurement
– Accuracy of the gas measurements
– Generation of alarms when defined alarm limits
are transgressed
If the gas measurement fails, an alarm is is-
sued.
WARNING
The following parameters are not monitored in
neonates: arrhythmia, cardiac output, and ST
segment analysis. No alarm will be activated.