Instructions for use Zeus Infinity Empowered SW 2.n 429
Technical data
Essential performance
The essential performance features comprise:
General
Supply of the anesthesia workstation with O
2
If the O2 supply (central gas supply or gas cyl-
inder) fails, an alarm is issued.
Supply of the patient with adequately oxygen-
ated breathing gas
If the breathing gas contains insufficient levels
of O
2, an alarm is issued.
Patients are not supplied with excessively high
anesthetic gas concentrations
If excessively high anesthetic gas concentra-
tions are delivered, an alarm is issued.
Monitoring of the airway pressure and the expir-
atory minute volume
Alarms are issued depending on the set alarm
limits.
Gas measurement
Accuracy of the gas measurements
Generation of alarms when defined alarm limits
are transgressed
If the gas measurement fails, an alarm is is-
sued.
WARNING
The following parameters are not monitored in
neonates: arrhythmia, cardiac output, and ST
segment analysis. No alarm will be activated.