Startup
92
Instructions for use Zeus Infinity Empowered SW 2.n
Checking the APL valve
(for manual checking in cases of emergency)
1 Adjust the pressure to 20 mbar (A).
2 Manually close the filter on the Y-piece or con-
nect the test lung.
3 Manually ventilate with breathing bag.
4 Check the pressure limits on the mechanical
pressure gauge (B).
Emergency ventilation bag
Example: Dräger Resutator 2000 (not shown).
z The emergency ventilation bag is present on
the device and its functionality has been
checked.
Checking the anesthetic gas scavenging
system AGSS
The anesthetic gas scavenging system (AGSS)
must conform to ISO 8835-3.
1 Transfer hose (A) is connected to the nozzle on
the back of the device and to the receiving sys-
tem (AGS).
2 The scavenging hose is connected to the re-
ceiving system, the plug (B) is in the Dräger wall
socket, which has a green indicator.
3 The float in the AGS hangs between the two
markings.
z Follow the instructions for use of the anesthetic
gas receiving systems (AGS) and the anes-
thetic gas scavenging system (AGSS).
215
A
B
WARNING
Risk of patient injury
A vacuum can be created in the breathing sys-
tem and in the patient's lung if the side open-
ings on the receiving system are covered.
Always make sure that the side openings on
the receiving system are not blocked.
044
A
B