Application
22
Instructions for use Zeus Infinity Empowered SW 2.n
Gas supply
The plug formats of the Zeus IE anesthesia work-
station are selectable and comply with the custom-
ary formats in the particular country.
Zeus IE has connections for O
2, Air or N2O backup
gas cylinders, which must be equipped with pres-
sure reducers and pressure sensors.
Gas disposal
The anesthetic gas receiving system (AGS) must
be used as a required accessory. The AGS is con-
nected to an appropriately marked connecting noz-
zle on Zeus IE.
Breathing system
Interfaces to the patient are the 22 mm ∅-sockets
on the breathing system for the connection of
EN 12342-compliant breathing hoses.
For breathing systems without rebreathing, e.g.,
Bain system, an external fresh-gas outlet is availa-
ble (optional).
Data exchange, ports
MEDIBUS is a software protocol for data transfer
between Zeus IE and external medical or non-med-
ical devices (e.g., hemodynamic monitors, data
management systems, or computers) via an RS232
interface.
Before transferring data, ensure strict compliance
with the following documents:
– MEDIBUS.X, Rules and Standards for Imple-
mentation (9052607)
– MEDIBUS.X, Profile Definition for Data Com-
munication V1.n (9052608)
– MEDIBUS, Supported Data SW 1.n (9040096)
Serial ports
Two serial ports, COM 1 and COM 2, are provided
for data communication using the Dräger MEDI-
BUS data protocol.
The COM 3 serial port can only be used in conjunc-
tion with the Integrated SmartPilot View option.
USB interface
The USB ports enable data exchange with Dräger
USB storage media.
Network interface
Zeus IE can be connected to a network using the
following interfaces.
– Infinity network connection in conjunction with
patient monitoring (optional).
– Network ports for remote diagnosis, printers,
and web applications, e.g., WebView (optional).
If a corresponding service contract has been
obtained, the function Remote Service can be
executed via a network connection and the hos-
pital network.
For further information on connections to IT net-
works, see page 436.