Instructions for use Zeus Infinity Empowered SW 2.n 95
Startup
Checklist
Use the checklist (A) to make sure the device is
ready for operation.
The checklist consists of the following test steps:
Supplies Safety (B)
Breathing system (C)
Suction unit Y-piece (D)
z Call up the test steps one after the other and
work through the checklist.
Supply & safety
1 Touch the Supplies/ Safety tab (B).
2 Perform the following test steps:
Power supply
Central gas supply
Gas cylinders
Emergency breathing bag
O
2 flush
O
2 emergency delivery
Breathing system
1 Touch the Breathing system tab (C).
2 Perform the following test steps:
Anesthetic gas scavenging system
Breathing system
Manual breathing bag
APL valve
Soda lime
Suction & Y-piece
1 Touch the Suction/ Y-piece tab (D).
2 Perform the following test steps:
Endotracheal suction
Microbial filters
Y-piece
Sample line
Watertraps
The checklist can also be called up during opera-
tion.
1 Switch to the system standby screen.
2 Touch the Device test button.
The Test details dialog window will be opened.
3 Touch the Checklist tab (A).
310
CAUTION
The checklist is part of the self test and must be
applied each time before operation. Otherwise
device operation may be impaired.
Tes t de tai l s
Self test
Self test
A
B
C
D