Instructions for use Zeus Infinity Empowered SW 2.n 411
Technical data
O2-, CO2 and anesthetic gas measurement
All measurements from side stream (sample gas
measurements) and under calibration conditions.
Sampling rate 200 mL/min ±20 mL/min
Delay time for sampling typically <4 sec. depending on sample gas line
Response time t
10…90 O2 <500 ms
Response time t10…90 CO2 <350 ms
Response time t10…90 anesthetic gases <500 ms
Warm-up phase after switching on Specified accuracy: typically after 14 min; max.
20 min
ISO accuracy: approx. 4 min
Dependence of accuracy on respiratory rate
(I:E ratio 1 : 1)
Specified accuracy: up to 40 1/min
ISO accuracy: up to 60 1/min
O
2 measurement (paramagnetic)
Measuring range: 0 to 100 Vol.%
Resolution: 1 Vol.%
Accuracy: ±(2.5 Vol.% + 2.5 % of the measured
value) in accordance with ISO 21647
CO
2measurement
Concentration in [Vol.%]: Measuring range: 0 to 10 Vol.%
Resolution: 0.1 Vol.%
Accuracy: ±(0.43 Vol.% + 8 % of the measured
value) in accordance with ISO 21647
Partial pressure in [mmHg]: Measuring range: 0 to 78 mmHg
Resolution: 1 mmHg
Accuracy: ±(3 mmHg + 8 % of the measured value)
in accordance with ISO 21647
Partial pressure in [kPa]: Measuring range: 0 to 10 kPa
Resolution: 0.1 kPa
Accuracy: ±(0.4 kPa + 8 % of the measured value)
in accordance with ISO 21647