Technical data
410
Instructions for use Zeus Infinity Empowered SW 2.n
Minute volume MV (spontaneous or mandatory) Measuring range: 0 to 40 L/min
Resolution: 0.1 L/min
Accuracy: ±10 % (under calibration conditions)
T
90 value on the display: approx. 45 s
Respiratory rate (RR or RRspon) Measuring range: 0 to 150 1/min
Resolution: 1 1/min
Accuracy: ±1 1/min
T
90 value on the display: approx. 45 s
Patient compliance Cdyn Measuring range: 0 to 300 mL/hPa
Resolution: 1 mL/hPa
Qualitative value only!
Average: 10 breaths
Patient resistance R Measuring range: 0 to 200 hPa/L/s
Resolution: 1 hPa/L/s
Qualitative value only!
Average: 10 breaths
O
2 flow tube Setting range: 0 to 16 L/min
Accuracy: ±10 % of the set value
Response time of the whole system
(according to ISO 80601-2-55)
4 +2 seconds
Measuring system (Cont.)