Instructions for use Zeus Infinity Empowered SW 2.n 13
For your safety and that of your patients
Information on electromagnetic compat-
ibility
General information on electromagnetic compatibil-
ity (EMC) according to international EMC standard
IEC 60601-1-2:
Medical electrical equipment is subject to special
precautionary measures concerning electromag-
netic compatibility (EMC) and must be installed and
put into operation in accordance with the provided
EMC information (see page 430).
Portable and mobile radio frequency communica-
tion equipment can affect medical electrical equip-
ment.
Sterile accessories
WARNING
Risk of electric shock
Do not connect connectors with an ESD warn-
ing symbol and do not touch their pins without
implementing ESD protective measures. Such
protective measures can include antistatic
clothing and shoes, touching a potential equi-
lization pin before and during connection of
the pins, or using electrically insulating and
antistatic gloves.
All relevant users must be instructed in these
ESD protective measures.
WARNING
Risk of device failure
Electromagnetic fields can compromise
proper operation of the device. Electromag-
netic fields are generated by, e.g., radio fre-
quency communication equipment such as:
– Mobile phones
– Radio frequency electrosurgical equip-
ment
– Defibrillators
– Shortwave therapy equipment
Use radio frequency equipment only with a
sufficient safety clearance of at least 20 cm
(7.9 in).
WARNING
Risk of electric shock
The connection of devices to auxiliary power
sockets can lead to an increased leakage cur-
rent. If the protective ground of one of these
devices fails, the leakage current may rise
above the permissible values.
– Only connect with the approval of the re-
spective device manufacturer.
– Have the leakage current checked by serv-
ice personnel.
– If the permissible value is exceeded, use a
mains power socket on a wall instead of
the auxiliary power socket of the device.
CAUTION
Risk of medical device failure and patient injury
Do not use accessories in sterile packaging if the
packaging has been opened, damaged, or if there
are other signs that the accessories are not ster-
ile.
Reuse, reprocessing, and sterilization of disposa-
ble products is not permitted.