Technical data
424
Instructions for use Zeus Infinity Empowered SW 2.n
SpO2 accuracy Measuring accuracy in Adult mode
(1)
:
SpO2:
0 to 69 % not specified
70 to 100 % sensor-specific, specified as follows:
Nellcor
(2, 3)
:
OxiMAX MAX-A, OxiMAX MAX-AL, Oxi-
MAX MAX-P, OxiMAX MAX-N, OxiMAX
MAX-I ±2
Nellcor:
DS100A ±3
Masimo
(2, 3)
:
LNOPADT, LNOPPED, LNOPNEO,
LNOPNEO SS, LNOP-YI ±2
Masimo:
LNOP-DCI, LNOP-DCIP, NR125 ±2
EAR ±3.5
Pulse:
±3 bpm or ±3 % (whichever is higher)
Measuring accuracy in Neonatal mode
(1, 2, 3, 4)
:
SpO2:
0 to 69 % not specified
70 to 100 % sensor-specific, specified as follows:
Nellcor:
N-25, OxiMAX MAX-N ±3
Masimo:
LNOPNEO, LNOPNEO SS, LNOP-YI ±3
Pulse:
±3 bpm or ±3 % (whichever is higher)
Pulse oximetry (SpO2) (Cont.)