Cleaning, disinfection, and sterilization
364
Instructions for use Zeus Infinity Empowered SW 2.n
Opening the breathing system
1 Open the five quick-release screws (A) with a
6 mm Allen key (supplied) by turning them 90°
counterclockwise.
2 Remove the upper plate.
3 Prepare the housing parts for cleaning and dis-
infection in the cleaning and disinfection ma-
chine.
Clean the APL valve and connecting nozzles sepa-
rately.
Removing the TurboVent 2 blower module
1 Using a suitable aid (e.g., a coin), open the
quick-release screws (A) by turning them 90°
counterclockwise.
2 Remove the blower module (B) from the breath-
ing system mount.
Perform a self test after sterilizing and changing the
blower module.
In conjunction with the TurboVent 2 MK05045
blower module, the IBF filter can be replaced by a
dust filter.
Device-end breathing system interface
Clean the interface to the breathing system using
wipe-down disinfection only.
043287
NOTE
Do not chemically disinfect/clean the blower mod-
ule and do not dismantle it.
A
A
B
WARNING
Only remove when the device is switched off!
Trained personnel only!
WARNING
If the absorber filter, HME filter, or the filters on
the device have not been installed, the blower
module may have become contaminated. It is
recommended that the blower module be re-
processed before changing the patient.
WARNING
Make sure that no liquids come into contact
with the openings of the measuring, suction
and delivery nozzles.
Make sure that no liquids get into the open-
ings of the ventilator or into the valve actua-
tors.