Technical data
406
Instructions for use Zeus Infinity Empowered SW 2.n
Pressure Support (Pressure support) Psupp
for Pressure Control, Volume Control AutoFlow
and CPAP
PEEP to 70 hPa
Accuracy: ±10 % or at least ±2 hPa
Resolution 1 hPa
CPAP MAN/SPON ventilation: 0 and 3 to 10 hPa
Accuracy: ±10 % or ±2 hPa
Resolution: 1 hPa
PEEP for Volume Control, Pressure Control, CPAP/Pres-
sure Support:
0 and 3 to 35 hPa; up to (P
max –5 hPa) or
(Pmax – Psupp)
Accuracy: ±10 % or ±2 hPa
Resolution: 1 hPa
Tidal volume (VT) Accuracy:
<100 mL: ±10 % or ±15 mL (under calibration con-
ditions)
>100 mL: ±10 % (under calibration conditions)
Resolution:
20 to 100 mL: 1 mL
100 to 500 mL: 5 mL
>500 mL: 10 mL
Respiratory rate (RR) 3 to 80 1/min
Accuracy: ±1 1/min
Resolution: 1 1/min
Inspiratory time T
i 0.2 to 10 seconds
Accuracy: ±0.1 seconds
Resolution: 0.1 seconds
Respiratory time ratio T
i:Te 4 : 1 to 1 : 4
Inspiratory pause ratio % Tplat:T
i with Volume
Control and Volume Control AutoFlow
20 to 50 %
Accuracy: ±20 %
Resolution 5 %
Rise time Slope for Pressure Control, Volume Con-
trol AutoFlow, and for Pressure Support breaths
0 to 2 seconds
Resolution: 0.1 seconds
Ventilation (Cont.)