Instructions for use Zeus Infinity Empowered SW 2.n 403
Technical data
O
2 concentration in the fresh gas,
Control mode FG ctrl.
25 to 100 Vol.% (measured under ATPD condi-
tions)
Accuracy depends on the carrier gas and the nom-
inal O2 concentration set:
O
2 set concentration
25 Vol.%
30 Vol.%
40 Vol.%
50 Vol.%
60 Vol.%
70 Vol.%
80 Vol.%
90 Vol.%
100 Vol.%
Resolution 1 Vol.%
Carrier gas
N
2O
±3.75 Vol.%
±4.2 Vol.%
±4.8 Vol.%
±5.0 Vol.%
±4.8 Vol.%
±4.2 Vol.%
±3.2 Vol.%
±1.8 Vol.%
0Vol.%
Carrier gas
Air
±0.8 Vol.%
±1.6 Vol.%
±2.9 Vol.%
±3.7 Vol.%
±4.0 Vol.%
±3.7 Vol.%
±3.0 Vol.%
±1.8 Vol.%
0Vol.%
Inspiratory O
2 concentration in the fresh gas, con-
trol mode Auto control
25 to 100 Vol.% (maximum achievable O2 concen-
tration because of 100 % relative air humidity in the
breathing system = 93 Vol.%. Corrected to BTPS
conditions)
Accuracy: ±(2.5 Vol.% + 2.5 % of the measured
value) in accordance with ISO 21647
Resolution: 1 Vol.%
ORC function Delivery of ≥250 mL/min O
2
>35 L/min
0 to 12 L/min
Accuracy: ±35 % or ±2 L/min
O
2 flush
O
2 safety flow
Central gas supply (Cont.)