Technical data
402
Instructions for use Zeus Infinity Empowered SW 2.n
Maximum gas withdrawal
O2 125 L/min
(including O2 flush, O2 emergency delivery and
maximum gas consumption at the O2 outlet)
Air 160 L/min
(including internal supply of the bronchial aspirator
and maximum gas consumption at the Air outlet)
N2O 20 L/min
Optional gas supply outlets The output pressure corresponds to the respective
input pressure
O2 outlet
(at maximum O2 gas delivery)
max. 41 L/min at 2.8 kPa x 100
Air outlet
(at maximum Air gas delivery) and
use, no external consumers
(e.g., suction))
max. 39 L/min at 2.8 kPa x 100
Driving gas consumption Air <1.5 L/min
Fresh-gas flow, control mode FG ctrl.: 0.25 to 18 L/min (STPD)
Accuracy ±10 %
Resolution 0.1 L/min
Fresh-gas flow, control mode Auto control: 0 to 18 L/min
Accuracy ±0.05 L/min or ±10 %
Resolution with flow
MIN = "Uptake": 0.015 L/min
otherwise
Resolution with flowMIN: 0.1 L/min
Central gas supply (Cont.)