Fault – Cause – Remedy
316
Instructions for use Zeus Infinity Empowered SW 2.n
Note 250 Vol. agent sensor
failure
Volatile agent sensor not op-
erable.
Continue anesthesia with
IV drug.
Call DrägerService.
Caution 210 Volume-controlled
vent. fail
Volume controlled ventilation
failed because of flow sensor
failure.
Use pressure controlled
ventilation.
Caution 060 Wrong volatile
agent detected
Detected agent is different
from selected type: – The
agent has been changed –
The DIVA module contains
the wrong agent
Check MAC values! Man.
delivery: flush with high
fresh gas flow Auto-deliv-
ery: system flushes auto-
matically! Exchange the
DIVA module, check con-
tent!