Instructions for use Zeus Infinity Empowered SW 2.n 315
Fault – Cause – Remedy
Warning 045 VA delivery cutoff DIVA failure. DIVA delivery
interrupted.
Continue anesthesia with
IV drug.
Call DrägerService.
Warning 045 VA delivery failure Failure of anesthetic agent
delivery system! Or delivery
terminated by protection sys-
tem. Delivery was switched
off!
Continue anesthesia with
IV drug.
Call DrägerService.
Warning 045 VA delivery off Failure of anesthetic agent
delivery.
Continue anesthesia with
IV drug.
Call DrägerService.
Warning 255 Vent+mixer+DIVA
cutoff
Failure of Ventilator, mixer
and DIVA. Mixer and DIVA
interrupted.
Open O
2 Safety valve.
Monitor ventilation by
press. gauge.
Call DrägerService.
Warning 255 Vent. fibr. / Vent.
tach.
Ventricular fibrillation: The
monitor identifies a sinusoi-
dal waveform with fibrillation
characteristics.
Check patient and take ap-
propriate action.
Note 100 Vent. turbine temp.
high
Temperature of ventilator tur-
bine is too high.
Reduce breathing pres-
sure! Ventilate manually!
Warning 135 Ventilator failure Failure of automatic ventila-
tion: - Ventilator turbine
failed - Temperature of venti-
lator too high.
Ventilate manually! Switch
to Fresh-gas control! Check
breathing system and tur-
bine. Call DrägerService.
Caution 000 Ventr. bigeminy Sequence of beats with the
pattern: normal, PVC, nor-
mal, PVC, normal
Check patient and take ap-
propriate action.
Caution 000 Ventr. couplet Sequence of beats with the
pattern: normal, PVC, PVC,
normal
Check patient and take ap-
propriate action.
Warning 000 Ventr. tachycardia Ventricular tachycardia: N or
more PVCs are detected in
an interval T = (60 * (N – 1))
/ R, where N = VT count and
R = VT-rate
Check patient and take ap-
propriate action.
Caution 000 Ventricular run Ventricular run: Series of 3 to
N-1 consecutive PVCs with a
beat-to-beat rate > the VT
rate.
Check patient and take ap-
propriate action.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy