Fault – Cause – Remedy
314
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 125 TOF count high TOF count is higher than the
the set alarm limit.
Check patient, treat if nec-
essary. Change alarm limit
as appropriate.
Caution 250 TOF count low TOF count is lower than the
set alarm limit.
Check patient, treat if nec-
essary. Change alarm limit
as appropriate.
Note 100 Too many pumps Maximum number of pumps
exceeded. System only sup-
ports up to 4 pumps.
Only 4 pumps are displayed
and can be controlled.
Note 100 Trident pod discon-
nected
Trident pod not connected to
system.
Check cables and connec-
tions. Replace cables, if
necessary.
Note 105 Trident pod hard-
ware fail
Trident pod hardware failure.
An unsupported pod is con-
nected.
Check Trident pod
connection. Disconnect
unsupported pod and
reconnect Trident pod. Call
DrägerService.
Note 150 Two BISx pods de-
tected
A second BISx Pod is con-
nected.
Remove the BISx pod that
was connected last.
Note 100 Two NMT pods de-
tected
A second NMT pod is con-
nected.
Remove the NMT pod that
was connected last.
Note 150 Two SpO
2 sensors
detected
A second SpO2 sensor is
connected.
Remove the SpO2 sensor
connected last.
Note 000 Unable to learn No dominant complex for
QRS processing found after
100 beats and until the end
of learning phase. Learning
will be continued.
Check electrode applica-
tion. Check leads. Choose
another lead for QRS
processing.
Note 020 V electrode fail Lead-off condition for indi-
cated lead detected: -Broken
cable – Loose lead wire
-Faulty lead wire – Wrong
lead -Dried out gel -Bad con-
tact.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Note 020 V+ electrode fail Lead-off condition for indi-
cated lead detected: -Broken
cable - Loose lead wire
-Faulty lead wire - Wrong
lead -Dried out gel -Bad con-
tact.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy