Instructions for use Zeus Infinity Empowered SW 2.n 313
Fault – Cause – Remedy
Note 050 Temp 1 unplugged Cable defective or un-
plugged.
Check equipment and re-
place if necessary.
Note 050 Temp 2 hardware
fail
Temperature measurement
system failure.
Call DrägerService.
Caution 200 Temp 2 high Temperature exceeds upper
alarm limit.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Caution 020 Temp 2 low Temperature exceeds lower
alarm limit.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Note 050 Temp 2 unplugged Cable defective or un-
plugged.
Check equipment and re-
place if necessary.
Caution 080 Tidal vol. high (P
min
ltd.)
At minimum airway pressure
tidal volume is higher then
set. – Breathing settings not
adequate. – Patient breathes
spontaneously.
Change ventilation mode or
VT settings.
Caution 080 Tidal volume high Pressure Control: Upper limit
of insp. tidal volume has
been exceeded multiple
times. Volume Control:
VTset x 1.3 exceeded.
Observe minute volume.
Check alarm limit of insp.
tidal volume and P
insp set-
ting.
Note 100 Tidal volume high Pressure Control: Upper limit
insp. tidal volume reached.
Volume Control:
VTset x 1.3 exceeded.
Check alarm limit of
insp.tidal volume. Switch off
AutoFlow mode.
Caution 045 Tidal volume not at-
tained
Due to fresh-gas deficiency
the set volume could not be
attained.
Increase fresh-gas flow,
seal leaks. Provide suffi-
cient ventilation.
Note 210 Time synchroniza-
tion
The clock has been synchro-
nized with system time.
This message will disap-
pear automatically after
15 seconds.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy