Fault – Cause – Remedy
312
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 125 Syringe 4 unlocked Infusion 4 can not start while
clasp is unlocked.
Lock clasp at the pump 4.
Control right position of the
syringe. and start infusion
or bolus.
Note 050 System restart re-
quired
System has been running for
more than 7 days.
Shut down and restart the
device at the next opportu-
nity.
Caution 050 System restart re-
quired
System has been running for
more than 14 days.
Shut down and restart the
device at the next opportu-
nity.
Warning 255 System restart re-
quired
Internal system error. Shut down and restart the
device as soon as possible.
Caution 150 T1 out of range Temperature value greater
or less than measuring
range
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Caution 150 T2 out of range Temperature value greater
or less than measuring
range
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Caution 000 Tachy Sinus tachycardia: N or more
consecutive normal beats,
with a beat-to-beat rate
>= TACH rate setting.
Check patient and take ap-
propriate action.
Note 050 Temp 1 hardware
fail
Temperature measurement
system failure.
Call DrägerService.
Caution 210 Temp 1 high Temperature exceeds upper
alarm limit.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Caution 020 Temp 1 low Temperature exceeds lower
alarm limit.
Check the patient and take
appropriate action. Check
equipment and replace if
necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy