Instructions for use Zeus Infinity Empowered SW 2.n 311
Fault – Cause – Remedy
Caution 050 Syringe 2 nearly
empty
Syringe 2 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Control syringe. Prepare
filled syringe, if necessary.
Caution 125 Syringe 2 unlocked Infusion 2 can not start while
clasp is unlocked.
Lock clasp at the pump 2.
Control right position of the
syringe. and start infusion
or bolus.
Warning 130 Syringe 3 empty Infusion 3 stopped.
Syringe 3 empty.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Caution 050 Syringe 3 nearly
empty
Syringe 3 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Control syringe. Prepare
filled syringe, if necessary.
Caution 125 Syringe 3 unlocked Infusion 3 can not start while
clasp is unlocked.
Lock clasp at the pump 3.
Control right position of the
syringe. and start infusion
or bolus.
Warning 130 Syringe 4 empty Infusion 4 stopped.
Syringe 4 empty.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Caution 050 Syringe 4 nearly
empty
Syringe 4 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Control syringe. Prepare
filled syringe, if necessary.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy