Fault – Cause – Remedy
310
Instructions for use Zeus Infinity Empowered SW 2.n
Caution 000 SV tachycardia Superventricular tachycar-
dia: N or more consecutive
normal beats, with a beat-to-
beat rate greater than or
equal to the SVT setting.
Check patient and take ap-
propriate action.
Note 050 Sync fail Defibrillator sync. output
failed
Call DrägerService.
Caution 210 Sync. ventilation
failure
Insp. and exp. flow sensor
failed. No triggering for as-
sisted ventilation available.
Observe sensor error mes-
sages. Use alternative ven-
tilator mode.
Warning 130 Syringe 1 empty Infusion 1 stopped.
Syringe 1 empty.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Caution 050 Syringe 1 nearly
empty
Syringe 1 nearly empty.
5 min until end of infusion or
10 % volume left. (depend-
ing on infusion rate and type
of syringe)
Control syringe. Prepare
filled syringe, if necessary.
Caution 125 Syringe 1 unlocked Infusion 1 can not start while
clasp is unlocked.
Lock clasp at the pump 1.
Control right position of the
syringe. and start infusion
or bolus.
Warning 130 Syringe 2 empty Infusion 2 stopped.
Syringe 2 empty.
Control syringe and patient.
Replace syringe, if neces-
sary. To drain syringe com-
pletely, confirm the green
button at the pump. (Only
possible, if 'Empty syringe
mode' is configured).
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy