Instructions for use Zeus Infinity Empowered SW 2.n 309
Fault – Cause – Remedy
Caution 200 ST I high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST I low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST II high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST II low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST III high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST III low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST V high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST V low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST V+ high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST V+ low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy