Fault – Cause – Remedy
308
Instructions for use Zeus Infinity Empowered SW 2.n
Note 200 SpO2 sensor un-
plugged
Sensor cable not connected
to the MultiMed pod.
Check to ensure that cables
are securely connected.
Check for defective sensor.
Note 010 SQI too low for BIS SQI (Signal Quality Index)
value goes below 15 %.
Check sensor. This may oc-
cur as a result of artefact
caused e.g. by eyeblinks or
motion. Isolate patient from
auxiliary equipment. Isolate
patient from extraneous
movement or vibration.
Check patient. Treat, if nec-
essary.
Caution 200 ST aVF high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST aVF low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST aVL high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST aVL low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST aVR high ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Caution 200 ST aVR low ST value is outside the upper
or lower alarm limit.
Observe the patient care-
fully and treat if clinically in-
dicated. Change the alarm
limits.
Note 160 ST cannot analyze The monitor cannot deter-
mine ST values owing to:
-Absence of normal beats
-Artefact
Perform a 'Relearn'. Calm
the patient. Check elec-
trodes and skin contact.
Isolate from auxiliary equip-
ment if possible.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy