Instructions for use Zeus Infinity Empowered SW 2.n 307
Fault – Cause – Remedy
Caution 250 Simulated patient
data
Patient data is simulated by
software.
Use simulated patient data
for demonstration purposes
only.
Warning 000 Sinus bradycardia 8 or more consecutive nor-
mal beats, with an average
rate < sinus bradycardia rate
setting. In neonatal mode:
heart rate lower BRDY-Set-
ting
Check patient and take ap-
propriate action. Check Ar-
rhythmia settings. Neona-
tal: Check BRADY settings.
Warning 000 Sinus bradycardia 8 or more consecutive nor-
mal beats, with an average
rate < sinus bradycardia rate
setting. In neonatal mode:
heart rate lower BRDY-Set-
ting
Check patient and take ap-
propriate action. Check Ar-
rhythmia settings. Neona-
tal: Check BRADY settings.
Warning 120 SmartCare can-
celed
Patient values are not within
the etCO
2 and VT ranges. A
technical problem occured.
Check patient condition.
Remedy technical alarms.
Check ventilation settings.
Note 100 Sound system er-
ror
The sound system failed.
There will be no audible
alarms.
Only use system in case of
emergency. Observe visual
messages at all times.
Caution 125 SpO
2 high SpO2 is above the current
upper alarm limit.
Observe the patient and
take appropriate action.
Note 100 SpO
2 HW failure SpO2 measurement system
hardware failure.
Call DrägerService.
Warning 250 SpO
2 low SpO2 is below the current
lower alarm limit.
Observe the patient and
take appropriate action.
Note 250 SpO
2 pulse search The monitor has not been
able to compute a valid
measurement within the last
30 seconds because of un-
stable measurement condi-
tions.
Ensure the SpO2 sensor is
properly attached to the pa-
tient and all cables are
properly connected. Call
DrägerService.
Note 150 SpO
2 sensor con-
nected?
Too much or insufficient light
is reaching the sensor or
measurement artefact Weak
signal due to poor perfusion
or low body temperature.
Check sensor attachment
Check patient's condition.
Check sensor and cables.
Relocate sensor to another
extremity.
Note 100 SpO
2 sensor
invalid
The monitor does not recog-
nize the connected sensor
as valid.
Check for defective or
unapproved sensor.
Replace the sensor. Call
DrägerService.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy