Fault – Cause – Remedy
306
Instructions for use Zeus Infinity Empowered SW 2.n
Warning 255 Safety settings ac-
tive
Due to MIXER + DIVA fail-
ure, safety settings have
been activated.
Check for additional mes-
sages. Check O2 and DIVA
settings.
Note 180 Sample line dis-
connected?
Sample line leakage! Check CO
2 sample line or
water trap -CO2-.
Warning 100 SC: Internal error A software runtime error has
occured. SmartCare is not
available.
A software reset is required
before SmartCare can be
restarted. Press 'Software
reset?' button. The system
will reboot and be available
again after approximately
30 seconds.
Warning 110 SC: Patient not
suitable
SmartCare cannot be contin-
ued. Patient values are not
within the etCO
2 and VT
ranges.
Check patient data. Check
patient condition. Check
ventilation settings.
Caution 095 SC: Recovery
failed
SmartCare could not induce
spontaneous breathing dur-
ing recovery.
Check patient medication.
Check settings for Trigger.
Wait until patient is ready
for recovery.
Note 195 SC: Respiratory
drive might be suffi-
cient
The patient might be ready
for extubation.
Check the respiratory drive
of the patient to determine if
the patient can be extu-
bated.
Note 195 SC: Respiratory
drive sufficient
The respiratory drive is suffi-
cient.
Consider extubating the pa-
tient.
Warning 100 SC: Severe hypov-
entilation
etCO
2 is too high. Check ventilation parame-
ters.
Caution 085 SC: Spont. breath-
ing detected
Spontaneous breathing oc-
curred during controlled ven-
tilation.
Check patient medication.
Allow spontaneous breath-
ing.
Note 195 SC: Spontaneous
breathing insuffi-
cient
Spontaneous breathing ac-
tivity of patient is insufficient.
Check patient medication.
Check settings for Trigger.
Caution 100 Service access al-
lowed
Warning: The service access
via cable connection is al-
lowed! Ensure that no pa-
tient is connected.
After finishing service deac-
tivate 'Service access al-
lowance'!
Caution 000 Setting not con-
firmed
Setting or change not con-
firmed within the timeout of
15 sec.
Repeat last setting or
change if needed.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy