Instructions for use Zeus Infinity Empowered SW 2.n 305
Fault – Cause – Remedy
Note 150 RV nonstatic Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Caution 050 RV out of range Invasive pressure signal out
of measurement range.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Caution 125 RV S high Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Caution 130 RV S low Invasive pressure value out-
side alarm limits.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Warning 250 RV static Static pressure -No pulse,
-Press. transducer turned
off, -Catheter tip lodged
against vessel wall, -Clot on
the catheter tip
Check patient and take ap-
propriate action. Open sys-
tem by turning the 3-way
tap. Follow procedure for
dislodging catheters. Fol-
low procedure for clotted
catheters.
Note 050 RV transducer fail Pressure transducer failed.
Invasive pressure cannot be
measured.
Check transducer and con-
nections. Exchange trans-
ducer.
Note 025 RV transducer un-
plugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 RV zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Caution 150 Safety O
2 opened Safety O2 valve opened. Ad-
ditional O2 fresh gas is deliv-
ered.
To be used in case of emer-
gency only. Observe O2-
and vol. agent concentra-
tion. At the end of operation
close the valve!
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy