Operation
120
Instructions for use Zeus Infinity Empowered SW 2.n
Default settings:
Parameter Setting range Default settings:
Neo Ped1 Ped2, Adult1,
Adult2
Carrier gas
Air or N
2O
1)
1) A nitrous oxide-free device system can be configured under Basic settings.
0 to 100 Vol.% Air
0 to 75 Vol.% N
2O
N
2ON2ON2O
FiO2 25 to 100 Vol.% 100 Vol.% 100 Vol.% 100 Vol.%
MinFG flow ; 0.25 to 18 L/min
Flush flow 3 to 18 L/min 4 L/min 6 L/min 6 L/min
etISO 0 to 5 Vol.% 0 Vol.% 0 Vol.% 0 Vol.%
inISOmax 0 to 5 Vol.% 4 Vol.% 4 Vol.% 4 Vol.%
etSEV 0 to 8 Vol.% 0 Vol.% 0 Vol.% 0 Vol.%
inSEVmax 0 to 8 Vol.% 8 Vol.% 8 Vol.% 8 Vol.%
etDES 0 to 18 Vol.% 0 Vol.% 0 Vol.% 0 Vol.%
inDESmax 0 to 18 Vol.% 12 Vol.% 12 Vol.% 18 Vol.%