Operation
150
Instructions for use Zeus Infinity Empowered SW 2.n
Basic setting
z Touch the Edit button (D).
The Drug editor dialog window will be opened.
z Touch the General settings tab (F).
The basic drug settings are displayed:
G Name
H Abbreviation
I Concentration
J Concentration unit: µg/mL, mg/mL
K Quantity unit: mL, mg, µg/kg, mg/kg
L Time unit: min, h
Setting dose
z Touch the Dose settings tab (M).
The infusion and bolus delivery settings are dis-
played:
N Infusion rate (mL/h)
O Maximum infusion rate (mL/h)
P Initial bolus (mL)
Q Maintenance bolus (mL)
R Bolus rate (mL/h)
S Maximum bolus rate (mL/h)
Units can be defined by the user and may vary.
426
NOTE
If a drug is stored in a profile, the abbreviation
cannot be changed. The drug must first be re-
moved from the profile.
Drug editor
F
G
H
I
J
K
L
427
M
N
O
P
Q
R
S
Drug editor