Instructions for use Zeus Infinity Empowered SW 2.n 157
Operation
2 Touch the Shut down button (D).
The system shuts down.
3 Close O
2 emergency delivery if necessary.
If the device is in system standby for longer than
90 minutes, the Zeus IE may need a warm-up
phase of 3 minutes before anesthetic agent deliv-
ery can resume.
When Zeus IE is not in use
z Leave the device connected to the mains sup-
ply. The green LED with the symbol sym-
bol lights up.
1 Close the backup gas cylinders (if used).
2 Disconnect supply hoses from the central sup-
ply, roll up and hang over the storage place on
the back of the device.
3 Pull the connector of the anesthetic gas scav-
enging system out of the wall socket.
4 Pull off the hose of the anesthetic gas scaveng-
ing system.
Without connection to the mains supply
307
WARNING
Remove gas supply hoses from the wall termi-
nal units (central supply).
If the gas supply hoses remain connected to
the wall terminal units, even minor internal
leaks may lead to contamination of the supply
gases.
NOTE
As the DIVA metering modules cool down slowly
(when using desflurane), metering accuracy is
temporarily reduced if Zeus IE is shut down and
put back into operation immediately.
Shut down?
D
NOTE
Charge the battery at least every 4 weeks or else
there will be no uninterruptible power supply avail-
able when the device is put into operation,
see page 58.