Monitoring
224
Instructions for use Zeus Infinity Empowered SW 2.n
During venostasis, the Stasis message, the cuff
pressure and the time remaining are displayed in
the parameter box.
Aborting venostasis:
z Touch the NIBP stop button.
After concluding the venostasis, the Single mode is
activated.
Inflation limits Neonates Pediatric patients Adults
Cuff pressure (mmHg) Neo. 140: 40 ±3 Ped. 180: 60 ±4 Adult 270: 80 ±5
Inflation time (s) 60 ±2.5 120 ±5 120 ±5