Instructions for use Zeus Infinity Empowered SW 2.n 227
Monitoring
Preparing pressure transducers
Zeroing
Pressure transducers must be zeroed under the fol-
lowing conditions:
– Immediately after inserting the catheter into the
circulatory system of the patient
– After connecting the pressure transducer to a
module
– Before starting monitor operation
– Before inputting a calibration factor
– After exchanging the breathing circuit or pres-
sure dome
Adjust static pressure transducers individually:
z Touch the Zero button in the Sensors/Parame-
ters > IBP > Cal. dialog window.
Zeroing several static pressure transducers simul-
taneously:
Prerequisite: The Zero all BPs button has been
configured, see page 332.
z Touch the Zero all BPs button.
Or
z Press SmartZero button on the hemody-
namic module (see page 202).
In the IBP curve, the Zeroing message appears
with a yellow background in the top right field. After
zeroing has been completed, the message fades
out again independent of whether the zeroing was
successful or not.