Instructions for use Zeus Infinity Empowered SW 2.n 261
Fault – Cause – Remedy
Note 025 ART transducer
unplugged
Pressure transducer for
specified parameter is either
unplugged or defective.
Active pressure: reconnect
or replace the cable. Inac-
tive pressure: turn off
alarms.
Note 050 ART zeroing Mercury calibration in
progress.
Complete calibration before
you begin monitoring the
patient.
Warning 255 Asystole No valid QRS complex de-
tected for the last 6 sec.
Check the patient's condi-
tion!
Warning 200 'Audio paused' key
stuck
Key is stuck or was pressed
for more than 10 seconds.
Press 'Audio paused' re-
peatedly. If alarm condition
persists, 'Audio paused'
cannot be used anymore.
Caution 075 Auto control cutoff Unable to operate in Auto
control mode. - Water trap? /
- Sample line blocked?
- sample line leak? - High
tube leak? - MixGas? /
- Soda lime exhausted?
Select mode: Fresh-gas
control
Note 255 Auto control cut-
off/high-flow
Unable to operate in Auto
control mode. - Water trap? /
-Sample line blocked? -
Sample line leak? - High
tube leak? - MixGas? / -Soda
lime exhausted?
Select mode: Fresh-gas
control!
Caution 255 AW press. protect.
fail
Airway pressure measure-
ment not plausible. Electr.
protection system for airway
pressure failed.
Check airway pressure
sensors. Observe pressure
gauge. Press 'Anesthesia
ventilation off' switch. Venti-
late manually. Perform self
test.
Warning 250 AW pressure high Upper limit of breathing pres-
sure or P
max exceeded.
Check settings for P
max and
Paw alarm limit. Check
spontaneous breathing.
Classifi-
cation
Pri-
ority
Alarm text Cause Remedy