Instructions for use Zeus Infinity Empowered SW 2.n 319
Fault – Cause – Remedy
Anesthetic agent flows out
when emptying with Quik Fil.
The bottle is not completely
screwed on or incorrectly
screwed on so the bottle valve
does not open.
Unscrew the bottle from the
emptying tool, then screw it on
again.
Escaping anesthetic agent
when filling/emptying the des-
flurane metering module
The O-ring on the anesthetic
agent bottle is damaged.
Remove bottle from the filling de-
vice. Perform the process again
with a new bottle.
Escaping anesthetic agent or
saturated vapor from opened
desflurane filling device (lock-
ing cover is removed from fill-
ing port).
Leaking valve or sealing com-
pound in the desflurane filling de-
vice.
Close the filling device with the
locking cover. Contact
DrägerService.
Remove bottle when filling before
it has been emptied completely.
Close the filling device with the
locking cover.