Maintenance
388
Instructions for use Zeus Infinity Empowered SW 2.n
Replacement by DrägerService or a technician specifically trained for this purpose
Lead-gel battery UPS (2 pcs.) After 3 years or the message Battery failure have it replaced by
trained personnel. Dispose of it in accordance with the local
waste disposal regulations.
Pressure controller (8417029) of the
gas supply block, (3 pcs.)
Replace after 6 years by DrägerService or a technician specifi-
cally trained for this purpose.
Sample gas return line filter Replace by trained personnel after 1 year.
Filters upstream of pressure sensors Replace by trained personnel after 1 year.
Filter mat of the power unit Replace by trained personnel after 1 year.
Diaphragms in the breathing system
(3 pcs.)
Replace after 2 years by trained personnel.
Sinter filter – gas supply block (3 pcs.) Replace after 2 years by DrägerService or a technician specifi-
cally trained for this purpose.
Dust filter – device ventilation Annual visual inspection, to be replaced by trained personnel af-
ter 3 years.
Central supply pressure reducers –
fresh-gas mixer (3 pcs.)
Replace after 6 years by trained personnel.
Replacement by trained personnel
Filter mat for gas measuring module Replace annually by trained personnel.
NBP air filter in the Apollo drawer Replace after 2 years by trained personnel.
O-rings (6870522) on water trap
holder, (2 pcs.)
Replace after 2 years by trained personnel.
O-ring (6870529) on water trap holder Replace after 2 years by trained personnel.
Small sealing rings (8602998) breath-
ing system, (6 pcs.)
Replace after 3 years by trained personnel.
Large sealing rings (M33729) breath-
ing system, (3 pcs.)
Replace after 3 years by trained personnel.
Guide sleeves (MK03310) DIVA me-
tering module, (2 pcs.)
Replace after 3 years by trained personnel.
Valve body (MK01580) DIVA meter-
ing module, (2 pcs.)
Replace after 3 years by trained personnel.
O-rings (MK02884) DIVA metering
module, (2 pcs.)
Replace after 3 years by trained personnel.
Distance washers (MK03284) DIVA
metering module, (2 pcs.)
Replace after 3 years by trained personnel.
Lithium batteries for the Diana and
Verona circuit boards
Replace after 6 years by trained personnel.
Lithium battery for the Dirona circuit
board (alternative)
Replace after 6 years by trained personnel.