Assembly and preparation
68
Instructions for use Zeus Infinity Empowered SW 2.n
Pulling out the used absorber
1 Press the unlock button (A): the holder flips
open.
2 Slide the disposable CO
2 absorber (C) out of
the holder (B).
3 Dispose of the used absorber.
Information about disposal is available in the in-
structions for use of the Drägersorb CLIC absorber.
Installing the new CO
2 absorber
1 Shake the disposable CO
2 absorber before in-
serting, e.g. by inverting it several times to
loosen the soda lime.
2 Remove the seal from the new disposable CO2
absorber.
3 Slide the new disposable CO
2 absorber (C) as
far as possible into the holder (B) until you hear
it snap into place in the CLIC adapter.
068
A
C